Brand:
Product Code: OM 46-0619
- +

8 1/2" (216mm)

maximum capacity: 1.5mm

17 gauge cerclage wire

Useful for cerclage work in deep and restricted areas.Disassembles easily forcleaning

How to Use the Loute Wire Tightener

  1. Loosen both knobs by turning counter-clockwise until center pin disappears from view.
  2. Pass wire around bone and feed both ends into center hole at the tip.
  3. Use Wire Pulling Forceps (OM 46-0637) to pull ends through and hold wire taut while pushing the instrument close to the bone.
  4. Turn lower(distal) knob clockwise to lock wire in place. Then turn upper (proximal) knob clockwise to pull in remaining slack from wire.
  5. Balance instrument in one hand and rotate it with the other hand until resistance is felt. Then hold tightener shaft tight and turn lower (distal) knob clockwise until wire is cut.Note: Hold instrument straight (do not tilt) while tightening.